(3/08) Calibration
-
Page 5-14
13.
The cotton sample beard(s) are discarded automatically.
14.
The results are displayed on the screen.
15.
After the second, fourth, and sixth tests a window launches asking
the operator to refresh the surface of the cotton sample and the
sampler door will open. Fluff and rotate the cotton in the sampler
drum(s) then press the
S
TART
button on the top of the cabinet.
16.
Complete the remainder of the tests using the steps above.
17.
Click
Finish
to return to the C
ALIBRATION
M
AIN
M
ENU
.
3.1.4.2
Long Cotton Calibration
1.
Be sure to check that the standard values for long/strong cotton
displayed on the screen are the same as the values on the box of
long/strong cotton you are using for calibration.
2.
Now perform Test 1 through 12 for long/strong cotton.
3.
Repeat the steps you used for testing short/weak cotton, but use
the long/strong calibration cotton, rotating and exchanging the
cottons in the same order.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......