USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 4
(3/08) Configuration
-
Page 4-12
All Offset Constants (except Trash Area) MUST remain at 0
unless you are instructed to change one of the constants by
an Uster service technician.
Trash area should be 1.0; it is a slope rather than an offset.
There is no selection to enable/disable offsets however an offset with a
value of zero (0) is not altered.
6
Diagnostics Log
This area enables the operator to configure topics that will be reported
into the log file. You can change these options between different testing
cycles or have default options for all cycles.
Place a check mark in the box preceding each desired selection and
select
Apply
to save.
Figure 4-9: Diagnostics Log
You can also check the
Save as the Default
box before selecting
Apply
.
The
Log Mic Scale Serial Input
checkbox is used to log serial input
from the balance. This is normally OFF, because the balance sends
data every second, whether testing is being done or not.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......