Chapter 5
(3/08) Calibration
-
Page 5-20
4.3.1
View Calibration Results
Figure 5-11: View Mic Calibration Results
To view the results:
1.
Click the
View Calibration Results
button (on the M
AIN
C
ALIBRATION
M
ENU
).
2.
Enter a valid password and click
OK
.
3.
Select
Cancel
to return to the C
ALIBRATION
M
AIN
M
ENU
.
5
Color and Trash Calibration
The calibration procedure for the color testing mechanism is very simple.
The system observes a set of tiles of known color values and compares
the values read from those tiles to the calibration constants stored in the
system. If the values do not agree, the system alters the constants to
bring the measurements back to known standards. The last tile
observed is the trash tile.
Standard Tile Values and Trash Tile Values should be entered before
performing Tile Calibration.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......