USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 8
(5/08) System
Diagnostics
-
Page 8-16
3.7.1
Tare Moisture Channels
Click
Move Combs to Home
or
Move Combs to L/S 2
to begin.
There are three selections that can be used to tare moisture channels:
1.
Tare Left
2.
Tare Right
3.
Tare Both
Note:
Before performing the Tare procedure, ensure that there is
NO cotton in the combs.
This area also allows you to perform the following actions:
4.
Move Combs to Drum
5.
Move Combs to L/S 1
6.
Move Combs to L/S 2
7.
Move Combs to Home
3.7.2
Moisture Channel Tests
This area allows you to do the following:
1.
Read Data
(returns measurements from both combs)
2.
Read Data by Closing Front Jaw
(automatically returns data)
3.
Ready Data by Opening Front Jaw
(returns information with jaw
open)
When testing is complete, click the
Finish
button to return to the
D
IAGNOSTICS
M
AIN
M
ENU
.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......