Chapter 8
(5/08) System
Diagnostics
-
Page 8-9
3.4.1
Mic Air Setup
Insert the micronaire calibration plug and use your finger to check the
Mic Air Reading in the following ways:
1.
Click
Mic Air Off
.
2.
Click
Measure Mic Tare
. Values will appear to the left of the
screen.
3.
Click
Mic Air On
.
4.
Block the orifice on the micronaire calibration plug. Mic Air
Reading should be 4.0 +/- 0.1.
5.
Block the orifice on the rear of the air bridge assembly. Mic Air
Reading should be -4.0 +/- 0.1.
6.
Block both the micronaire calibration plug orifice and the rear air
bridge assembly orifice. Mic Air Reading should be ~-0.00 +/-
0.05.
7.
Click
Mic Air Off
.
8.
Click
Finish
to return to the D
IAGNOSTICS
M
AIN
M
ENU
.
3.4.2
Mic Chamber Setup
To setup the micronaire chamber, do the following:
1.
Make sure that Mic Slope and Offset are set to their default
values.
2.
Insert 10.0 grams of
High
Mic cotton into the micronaire
chamber.
3.
Close the Mic door.
4.
Adjust the micronaire chamber until the raw Mic value equals the
high Mic value.
5.
Click the
Finish
button to return to the D
IAGNOSTICS
M
AIN
M
ENU
.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......