USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 5
(3/08) Calibration
-
Page 5-7
3
Length/Strength Calibration
The length/strength module has one (1) calibration type available:
1.
LS Cotton
This can be selected to
Check Calibration
. The L/S Cotton
calibration type allows the operator to:
♦
Setup Tolerances
♦
Setup References
♦
View Calibration Results
3.1
L/S Cotton Calibration
1.
Highlight Strength in the Module window.
2.
Highlight
LS Cotton
in the Calibration Type window.
3.1.1
Setup L/S Calibration Tolerances
Range tolerances ensure that the slope is not distorted by data that are
at opposite tolerance limits. For example, if the measured value for one
of the calibration cottons is at the upper limit of the tolerance, the range
value ensures that the corresponding value for the other calibration
cotton cannot be at the lower limit of the tolerance. A typical range
tolerance is 1.5 times the corresponding tolerance.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......