USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 5
(3/08) Calibration
-
Page 5-28
5.2.3
Check Calibration
Highlight
Color + Trash
in the Module Window, then highlight
Trash
Tiles
in the Calibration window and click
Check Calibration
. This
will launch the Trash Tile Calibration screen and a trash image window
from which you can select
Show Trash
or
Show Image
.
Figure 5-19: Trash Tile Calibration Screen
The Status window displays instructions for completing calibration.
1.
Click the
S
TART
button
(on the screen)
to begin calibration.
2.
Hold the
WHITE
tile glossy side down to the window on the color
head and press the
Start
button
(on the Mic cabinet)
.
The push
hand will
NOT
operate during calibration.
3.
Hold the
TRASH
calibration tile (approximately 200 dots) onto the
color head window and press the
Start
button
(on the Mic
cabinet)
.
4.
The system will prompt you
(several times)
to reposition the
TRASH
tile and click
OK
on the screen as it progresses through calibration
for Trash Rectangle Size and for Trash Area.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......