USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 6
(3/08) System
Testing
-
Page 6-23
10.
Weigh a +/-10 gram cotton sample on the balance.
11.
Place the cotton sample
(of configured valid range)
in the
micronaire chamber and close the door.
12.
The test is performed and the sample ejected.
13.
The number of times a sample is tested is determined by the
selections made in Configuration
;
the test results displayed on the
screen are an average of the observations.
As the tests are completed for the cotton sample, the results are
displayed on the screen and transmitted to the host
(if testing online)
.
8.2
Retesting
Once data is collected and the summary of data is displayed, a bale
with discrepant data (or data that appears to be incorrect) can be
retested by re-entering the Bale ID and repeating the testing cycle. Initial
information from a retested bale will be overwritten when transmitted to
the host, or when it becomes part of the testing report.
Discrepant data includes lot limit results. Lot Limits are indicated by a
BLUE
background.
Note:
Discrepant means the bale has had one (1) Lot Limit result.
If the Length/Strength, Color, or Trash test results indicate that the
measurements from the two sides of the bale are not within the tolerance
limits, an error message is displayed beside the measured value for the
field out of tolerance. A window will launch and ask if you want to
Accept
or
Retest
.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......