Chapter 8
(5/08) System
Diagnostics
-
Page 8-30
3.9.6.1
L/S Step by Step Test
1.
Place cotton sample into the sampler drums.
2.
Click
Initialize System
to move the combs into position in front
of the sampler drums.
3.
Click
Load Sample
to allow the beard to be collected.
4.
Click
Move to Brush
to collect the beard and move the combs
to the brush position.
5.
Click
Brush Beard 1
to brush the beard from the combs.
Click each remaining button in order (left column, then right column) top
to bottom to complete the Fibrogram/Length tests.
3.9.6.2
Exit
When testing is complete
Finish
to return to the D
IAGNOSTICS
M
AIN
M
ENU
.