USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 5
(3/08) Calibration
-
Page 5-23
5.1.3
Check Calibration
Highlight
Color + Trash
in the Module Window, then highlight
Color
Tiles
in the Calibration window and click
Check Calibration
. This
will launch the Color Tiles Calibration screen and a trash image window
from which you can select
Show Trash
or
Show Image
.
Figure 5-15: Color Tile Calibration Screen
The Status window displays instructions for completing calibration.
1.
Click the
S
TART
button
(on the screen)
to begin calibration.
2.
Place the
WHITE
tile face down on the color head and press the
Start
button
(on the Mic cabinet)
.
3.
Place the
BROWN
tile face down on the color head and press the
Start
button
(on the Mic cabinet)
.
4.
Place the YELLOW tile face down on the color head and press the
Start
button
(on the Mic cabinet)
.
5.
Place the
GRAY
tile face down on the color head and press the
Start
button
(on the Mic cabinet)
.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......