USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 8
(5/08) System
Diagnostics
-
Page 8-26
3.9.4.3
Length Motor Position Tests
To test the positions you set previously, click the following buttons:
1.
Move Breaker to Home
2.
Move Breaker to Comb 1 Front
3.
Move Breaker to Comb 2 Front
Observe the movements and repeat positioning as necessary.
3.9.4.4
Length Motor Current Check
This area is used to check the alignment and movement of the breaker.
The current will increase if the breaker is moving erratically, or making
noise during measurement. This procedure should only be performed
by a qualified service technician.
3.9.4.5
Save and Exit
When testing and setup is complete, click
Save Positions
. Once
positions have been saved, click
Finish
to return to the D
IAGNOSTICS
M
AIN
M
ENU
.