USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 5
(3/08) Calibration
-
Page 5-3
2.
For each repetition performed:
a.
If the data is invalid, an error message is displayed, and the rep
can be rejected.
b.
If the data is valid, it is accepted and displayed. The statistics
bar is updated with the Mean, Std dev., and %CV value for each
data item, for LS calibration only.
c.
In general there is no Bale ID dialog box. Calibration starts
when the user clicks the
Start
button on the screen. The
exception is trash calibration: it has a
Start
button to explicitly
force acceptance of the calibration options before measurement
is allowed.
d.
The
Report
button is enabled at each repetition count for report
output.
e.
Because LS and CT module calibrations do not support an odd
number of repetitions, an invalid measurement on one comb or
on the color head invalidates both measurements.
3.
When the calibration process is completed, there is no mechanism
for re-initiating calibration. The operator must exit calibration and
start
Check Calibration
again.
4.
Mic calibration does not support a calibration check. New
calibration constants are generated each time the procedure is
completed. The same is true for trash area deflation calibration.
5.
The calibration procedure for LS, color, and trash support a “check
first” mode, where collected data is compared to the reference to
determine if the existing calibration results yield final data that are
within tolerances, or not.
a.
For LS, checking is performed at the end of each reference set,
short + weak and long + strong. For color, the check is
performed when data collection is complete. If the collected data
pass tolerances, the overall check succeeds, no new calibration
constants are needed, and the saving of new calibration results is
disabled. If the checking process fails, the user is notified and
the procedure display indicates re-calibration is occurring. At the
end of data collection, the system determines whether the new
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......