Length/Strength Control Board
Error
Code
Message
Action
32/35
Brush Motor Fails To Move.
33/36
Brush Motor Position Error.
Go to L/S CTRL Board
Diagnostics and turn the Brush
Motor ON/OFF using the
buttons in the Digital Inputs
column. Notice if sensors turn
ON/OFF (above). If functional,
go to Brush Motor Diagnostics
and Learn Positions/perform
position tests. May indicate a
bad motor.
34/37
Brush Motor Temperature Too High.
Possible Motor Defect.
38
Drum Motor State Sensor Error.
1. Check for loose cable at the
Length Motor and at the
power box.
2. Go to L/S CTRL Board
Diagnostics and turn the
Drum Motor ON/OFF using
the buttons in the Digital
Inputs column. Notice if
sensors turn ON/OFF
(above). If sensors
unresponsive, may be a
cable, sensor or control
board problem.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......