Chapter 5
(3/08) Calibration
-
Page 5-4
calibration constants yield final data that pass calibration
tolerances. The status of this check is displayed. If Pass, then the
calibration data is saved automatically.
b.
Note that the calibration procedure for color supports the
checking of trash calibration (since the trash tile is part of the 6
tile color calibration / check set). The status of trash calibration
for the color head is displayed; however, to re-calibrate trash, the
user must close color calibration, open trash calibration, and then
use the separate trash calibration tiles to calibrate trash.