USTER®
HVI 1000
Instruction Manual
Uster Technologies, Inc.
Chapter 8
(5/08) System
Diagnostics
-
Page 8-3
3
Diagnostics/Hardware Setup
This set of diagnostics allows a qualified service technician to
troubleshoot individual motors of the instrument, to test the overall
operation of the Motor Controllers, Optics and Air system, and to
update the firmware for the Micronaire and the Length/Strength
cabinets.
3.1
BaleID BCR
This area is used for troubleshooting the bar code reader.
Figure 8-3: Bale ID BCR Screen
To test the bar code reader:
1.
Read Bar code
(use the bar code reader to scan the bale ID).
2. Type in a
Command
(see the bar code reader manual for a list of
commands).
3. Click
Send
(to send the command to the bar code reader).
4. Click
Init BCR
(to perform a default bar code reader initialization).
5. Click
OK
when finished.
Содержание HVI1000
Страница 2: ......
Страница 24: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 1 3 08 Introduction Page 1 10 ...
Страница 50: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 3 3 08 System Startup Page 3 16 ...
Страница 174: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 8 5 08 System Diagnostics Page 8 38 ...
Страница 192: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 10 3 08 Maintenance Page 10 6 ...
Страница 230: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 12 7 07 Appendix Page 12 30 ...
Страница 236: ...USTER HVI 1000 Instruction Manual Uster Technologies Inc Chapter 13 3 08 Index Page 13 6 W Warm Up Color Head 8 6 ...
Страница 238: ......