Page 78
CAUTION!
Under no circumstance should water be filled above the safety
valve holder.
In case more water is accidentally filled above the blue area or safety
valve, decrease the water level by draining the reservoir before starting
a cycle (see sec. 10.4).
Note:
This reservoir is designed with an overflow and filling the reservoir
above the safe level as indicated on the Front Fill Funnel will cause
excess water to spill out below the machine onto the counter.
CAUTION!
USE STEAM DISTILLED WATER ONLY having the characteristics
as per table in sec. 1.22.
The impurities in water from a well, spring, or
municipal water supply will create the need for more frequent cleaning
and maintenance.
This type of water can damage your instruments, your unit and void
your warranty.
Caution!
Wastewater should be brought into the public net in
accordance with the local rules or requirements
ONLY NON-HAZARDOUS LIQUIDS SHALL BE DISPOSED IN
PUBLIC SEWAGE!
7
6