The Elara11 offers a choice of the following automatic programs
designed to match the material to be sterilized.
Program # Program Name
Sterilization
Drying
Time
Temp
Time
1
Unwrapped Instruments
273°F
4 min
1 min
2
Wrapped Instruments
273°F
4 min
20 min
3
Unwrapped Delicate Instruments
250°F
20
min
1 min
4
Wrapped Delicate Instruments
250°F
20
min
20 min
Test 1
Vacuum Test
N/A
Pre-set
Test2
Bowie and Dick
273°F
3.5
min
2 min
The Elara11 is validated for use in sterilizing lumened devices of no
longer than 230mm and no smaller than 3.4mm for sterilization
programs 2 and 4.
The Elara11 is validated for use in sterilizing fabric packs/textiles for
sterilization programs 2 and 4.
The Elara11 is validated for use in sterilizing up to six dental
handpieces for sterilization programs 2 and 4.
A computerized control unit ensures a fully automatic sterilization cycle,
control and monitoring of physical parameters and clear documentation
of the sterilization cycle.
The Elara11 features a digital absolute pressure display which is used
for monitoring and control purposes. The device can display the
pressure in psia, psig, kPa, Bar A or Bar G
according to the operator’s
requirements. When the pressure is displayed in psig or Bar G, the
atmospheric pressure is shown (at sea level) as 0 psig. If the pressure
is defined in psia, Bar A
or kPa the absolute zero is displayed as “0”
and the atmospheric pressure is shown (at sea level) as 14.7 psia, 1
Bar A or 100 kPa respectively.
Note:
This unit comes from the factory with the pressure parameter set to
display in psig and the vacuum in inHg.
The Elara11 can display the temperature in ºF or ºC.
Note:
The unit comes from the factory set to display temperature in ºF.
The control system is designed to meet the most current sterilization
standards to ensure efficacy, safety of personnel and reliable
operation.