Page 42
5.
CHECKING AND CHANGING PARAMETERS AND OTHER DATA
This section shows how to access system data and modify parameters.
The Cycle Parameters directory containing parameters for controlling
the sterilization process is locked for programs 1 thru 4 and not
available for modification from the default values (except for drying).
Program 5 is a calibration program for use by a technician and
all parameters are locked. Program 6 is an extra drying program for
use by the operator if the load was not sufficiently dried. This program
is open for adding additional drying time.
Programs 7 & 8 are test programs, and all parameters are locked.
Spore testing is your only assurance of complete sterilization.
Once entering the programming mode, the operator will see and have
access to the following directory items.
Directory
Subdirectory
Quick Options see sec. 5.1
Add extra dry time
Export to USB
Print cycles
Version information
Start cycle by clock
Set date and time
Login
Exit
Main Menu
(requires
login) see
sec. 5.2
Cycle
Parameters
–
applicable only
for Custom
programs
(except Dry
Time) See sec.
5.2.1
Cycle Parameters
System
Parameters See
sec. 5.2.2
Print Rate All
Print Rate Sterilization
Screen Saver
Cycle Print Gap
Auto Wake Up
Maintenance
See sec. 5.2.3
Reset atmospheric pressure
Printer test
Print all gain and offset
Advanced
Options See
sec. 5.2.4
Set Temperature units
Set Pressure units