9.2.
Wastewater Reservoir
The Elara11 has two reservoirs, a mineral free reservoir for clean steam
distilled water and a wastewater reservoir for water that has gone through the
sterilization process. When the wastewater reservoir is full, the symbol
is
displayed. This situation is normal; however, the operator cannot run a new
cycle before draining the waste water reservoir (see10.4). The Elara11 comes
with an optional rear drain that can be connected to a building drain. If the
rear drain is connected, the wastewater reservoir will always be empty.
NOTE:
The temperature of the wastewater can be over 200°F (93°C). Plastic
drainpipe cannot tolerate temperatures above 140°F (60°C). An
external heat exchanger may be required to cool the discharge.
9.3.
Opening the Door
This machine is equipped with an electronic door lock. The door will not
open when the sterilizer is running a cycle, or the power is off, or an
error message is displayed.
Open the door by following these steps:
1.
Place your thumb on the plastic door cover (1) and the other fingers in
the handle (3).
2.
Pull the handle (2) until the latch of the door is released.
3.
Open the door.
4.
With the door open you can now fill the reservoir as per the
instructions in section (see 7.6).
Note:
The first time you start the unit it is important to set the date and time
(see sec. 5.1.6)