Page 77
7.6.
Filling the Mineral-Free Water Reservoir.
Use only steam distilled water having the characteristics
described in sec. 1.22
1. Open the door (1). The autoclave needs to be on for the door to open.
2. Pour steam distilled water, gently, into the front funnel (2) until it
reaches the top of the blue area (3) on the level gauge. If water is filled
above the blue area into the red area (4) then use the drain hose to
drain off the over fill (see sec. 10.4). It is preferable to use a carafe
or small pitcher when filling.
Note:
This reservoir is designed with an overflow and filling the reservoir
above the safe level as indicated on the Front Fill Funnel will cause
excess water to spill out below the machine onto the counter.
CAUTION!
Under no circumstance should water be filled higher than the blue
area (3) on the level gauge.
If the reservoir is empty, it can be filled quickly by adding water
directly through the opening at the top of the reservoir, as follows:
4.
Remove the water reservoir cover (6).
5. Pour steam distilled water into the reservoir through the opening on
top of the autoclave until it reaches the base of the safety valve
holder (7) or reaches the top of the blue area of the level gauge.
4
3
2
1