Page 92
10.3.
Periodically By the Operator
Caution!
Make sure the autoclave is not hot before cleaning it
1.
Once a month, check both safety valves.
2.
Once a month clean the strainer as per para. 10.6. Cleaning frequency
may be reduced according to experience.
3.
Replace the air filter, every 6 months or after 1000 cycles (whichever
comes first) according to 10.5.
4.
Every 3 months check the door gasket for any signs of physical
damage replace it if there is a tear or leakage.
10.4.
Draining the Reservoir
(Applies to the clean-water reservoir and to the waste-water reservoir)
The drain valve is located on the front right side of the autoclave after
the door is opened (1). The function of the drain valve is to drain the
water reservoir.
1.
To drain the reservoir, use item (5) with the plastic hose (6) attached to
it (supplied with the autoclave).
2.
Insert part (5) into valve (3) and press it until you hear a click. The
drain valve opens immediately, drain into a bucket.
3.
When the water reservoir is empty, press part (4). Item (5) will pop out
approx. 3mm and the drain valve will be closed. Remove item (5) with
the plastic tube.
4.
If the drained reservoir is the clean-water reservoir, fill reservoir with
distilled water until it reaches the full level. (Approximately 1.72gal (6.5
liters)).
5.
Turn on the main power switch.
6.
The autoclave is now ready for use.