Page 81
18.
Empty c a n i s t e r s should be placed upside-down, to prevent the
accumulation of water (see the figure below).
19.
When sterilizing glassware use only heat-proof glass. Glassware
needs to be placed on the tray with the open end down.
20.
Tubing should be rinsed after cleaning. When placed in the tray, make
sure that both ends of the tubing are open and there are no sharp
bends or twists.
21.
Cassettes or packs should be placed on the tray rack in place of the
trays. They should not be touching each other or the Chamber walls.
There should be about
½” (1.25cm) between cassettes or packs for
proper steam circulation.
22.
Cassettes in an ELARA11 should be sterilized in a vertical position
(see the figure below). To adjust the rack for vertical sterilization of
cassettes remove the trays and gently squeeze the sides of the rack
inward and at the same time rotate the rack into a vertical position.
Wrong
Right