Page 25
2.6.
Displayed Operational Messages / Symbols
Message /
Symbol Name
Message / Symbol
Description
Required Action
This symbol is displayed
when the door is open.
Close the door.
"Door is open"
This message is
displayed in stand-by
when the door is
opened, and
START/STOP is
pressed.
Close the door to perform a new
cycle. If the problem persists, call
for service.
"Cycle Ended"
This message is
displayed when the
cycle has ended
successfully.
Open the door. The Instruments
are ready to be removed.
"Test Ended"
This message is
displayed when the test
cycle has ended
successfully.
Open the door and select the
next cycle to run.
This symbol is displayed
when Cycle by Clock
mode is active.
Enter the Quick Options menu as
described in this manual to
change the time or to cancel this
option.
"Start cycle by
clock is active"
This message is
displayed if the user
presses START/STOP
key while the "start
cycle by clock" mode is
active. Starting another
cycle is not allowed.
Enter the Quick Options menu as
described in this manual to
change the time or to cancel this
option.
"Atmospheric
pressure not
set"
This message is
displayed when the
ATM needs to be set.
Turn the autoclave off then on.
Open the door for 2 minutes this
will automatically allow the
Atmospheric pressure to be set.
The chamber temperature must
be below 113°F.
This symbol is
displayed, and the cycle
will not start if the
temperature in the
jacket has not reached
the preset value.
Wait until the temperature in the
jacket reaches the preset value
This symbol is displayed
when there is no water
in the mineral-free water
reservoir.
Pour water in the front funnel until
it reaches the full level.