Page 104
Problem/
Error
Message
Message / Symbol Description
Corrective Action
"RTC Error -
Please Set
Current Date
and Time"
This message is displayed to set
the date and the time.
Set Current Date and
Time.
If the problem persists,
call the technician.
"Time Error"
The Date and Time have not been
set.
Set current Date And
Time. See sec. 5.1.6
This message is displayed if the
real time clock is faulty.
Call for service.
"Door is open
(During the
cycle)"
This message is displayed during
the cycle if the door is open, or the
door switch is faulty.
Close the door more
forcefully and perform a
new cycle. If the problem
persists, call for service.
"Canceled By
User"
This message is displayed any
time the START/STOP key is
pressed after the cycle has started;
the cycle will be aborted.
Wait until "cycle failed
–
canceled by user" or
"cycle end
– canceled by
user" is displayed. Press
the START/STOP key to
clear the error message
and perform a new cycle.
"Cycle Failed"
This message and symbol are
displayed if an error occurs before
sterilization cycle is completed.
Press the START/STOP
key to clear the error
message then perform a
new cycle. If the problem
persists, call for service.
"Test Failed"
This message and symbol are
displayed if an error occurs before
test cycle is completed.
Press the START/STOP
key to clear the error
message then perform a
new cycle. If the problem
persists, call for service.
"Air Error"
This message is displayed at the
end of the cycle if the autoclave
does not reach the atmospheric
pressure after 10 minutes.
Wait until the autoclave
reaches the atmospheric
pressure. Reset the
atmospheric pressure. If
the problem does not
correct itself call for
service
“Jacket is
cool”
The message is displayed if, when
pressing Start/Stop, the
temperature of the jacket is below
the preset temperature.
Wait until the jacket
warms up.
"Periodical
check time
exceeded -
Please call for
service"
The periodical maintenance time
has passed.
Call for service.