Page 64
without changing the parameter.
The custom programs are not FDA cleared and it is the use
r’s
responsibility to validate any custom cycles.
Spore testing is your only assurance of complete
sterilization.
5.3.2 Table of Parameters
This option is only available for a Custom program.
This table describes the function of each of the parameters that control
the different parts of the sterilization cycle. Each parameter is available
for modification by the user.
The program to be modified must be selected before entering the
"
MAIN MENU
" directory
Custom programs are not FDA cleared
.
It is the user’s responsibility to validate any Custom cycles.
This device utilizes vacuum pulses for removing air from the chamber
prior to sterilization. The Create Pulse parameters are what control this
process. Inappropriate parameter modification can affect the quality of
the sterilization process and the ability of the machine to complete that
process.
Spore testing is your only assurance of complete
sterilization.
Parameters
S
ubdire
c
to
ry
Description
Units
Range
Res
olutio
n
Purge time
Time for the purge
stage to last since the
purge temperature
has been achieved.
Purge is a stage when
steam is introduced
into the chamber and
exhaust is open to
allow all residual air to
be forced out by the
Minutes