Page 87
•
The program can only be changed when the door is open.
2. If needed additional drying time can be added at this time (see
Sec. 5.1.1).
Note:
Pushing the bottom tray all the way back will cause it to interfere with
the chamber sensors.
3. Close the door by either:
•
Holding the handle in the open position while pushing the door until it
comes to the closed position, then releasing the handle.
•
Pushing on the door handle and gently pushing the door closed.
•
When the door is properly closed and the mineral free reservoir is full
and the pre-heating is complete, the open-door symbol
is then
replaced by the message “System Ready”.
5.
Start the cycle by pressing the
START/STOP
key.
•
The door is now locked.
•
Stopping the cycle now will generate a cycle fail message.
9.6.
Cycle Description
The autoclave starts performing the sequence of operations.
The actual measured values of pressure and temperature are
displayed continuously and printed o n t h e p r i n t e r .
The display shows the current stage of the cycle. (See sec. 3.1).
In any program that has a drying stage scheduled, the dry stage
begins after the steam exhaust stage. The autoclave is equipped with
a vacuum pump that during the drying stage creates a vacuum in the
chamber drawing moisture from the instruments. Drying is performed
with the door closed.
At the end of a successful cycle, the screen shows the Cycle Ended
message and the door is automatically unlocked. The air valve is
o p e n e d to prevent formation of a vacuum.
In the event of a program failure, the exhaust valve is opened to
release pressure from the chamber and a fail message is displayed
There is a mandatory 1 minute of drying before the door can be
opened. When the mandatory drying is completed pressing
the
START/STOP
key will clear the error message and unlock the
door.