Page 71
End Temperature: 248°F Temperature when the cycle will end.
Dry Time: 1 min Dry time for the selected program
Ster. Time: 4.0 min Sterilization time for the selected
program
Ster. Temp: 273.2°F Sterilization temperature for the selected
program
Unwrapped Instruments Cycle name
Cycle Num:
000001
Cycle number
Software vers:
2.0.3.97
Software version
Version: 0
Model:
Elara11
Model name
Ser. Num:
000000000001
Model Serial number
Time:
08:33:29
Time of starting cycle
Date:
9/FEB/2010
Date of starting cycle
-------------------------------------------------------------
I/O Card vers:
7.3
Software vers:
2.0.3.97
Time:
08:33:29
Time of turning on
Date:
9/FEB/2010
Date of turning on
POWER ON
The device is turned on
Time:
00:00:00
Time of turning off
Date:
9/FEB/2010
Date of turning off
POWER OFF
The device is turned off
Legend
A
Vacuum stage (Pulse L & Pulse H)
D
Drying stage
H
Heating stage
S
Sterilization stage
CLK 1
Real Time Clock (see the explanation below)
CLK 2
Software clock (see the explanation below)
E
Exhaust stage
CLK1 and CLK2
The autoclave has two clocks on its main board: real-time clock and
processor clock. When sterilization starts, the system takes the
readings of both clocks and compares them (see the printout). At the
end of the sterilization stage, the discrepancy must not exceed 3
seconds.
6.2.
Printer Handling
6.2.1. Maintenance
Wipe off any soiling on the printer surface with a soft cloth and a weak
neutral detergent. After that, wipe the printer with a dry cloth.