Page 98
1.
Operate the sterilization cycle according to the manual, but with no
instruments.
2.
Allow a pressure of approximately 29 Psig (300KPA) to build up in the
chamber.
3.
Remove water reservoir cover (1).
CAUTION!
The next step will expose you to hot steam. To avoid being
burned by hot steam, keep all body parts away from the steam
flow.
1.
Pull the ring of each safety valve in turn using a tool, i.e., screwdriver,
hook etc., open the safety valve ring for 2 seconds, and then release it.
Be careful not to burn your hands.
2.
After 2-3 seconds the safety valve should close automatically.
3.
Press the
START/STOP
key to abort the cycle and allow the steam to
exhaust from chamber.
4.
Wait until the pressure goes down to zero, only then can the door be
opened.
5.
If the safety valve closes properly the procedure was successful.
6.
If the safety valve does not close or leaks steam or water when it
does close the safety valve needs to be replaced. This will require
an authorized service technician.
No.
Description
1
water reservoir cover
2
Safety valve
3
Pulling device
4
Pressure relief ring