10.6.
Replacing the Door Gasket
To avoid injuries, replace the gasket while the autoclave and
autoclave door are cold.
Pull off the gasket from the door groove and install the new gasket
referring to the directions below.
1.
Pull off the gasket from the groove.
2.
Clean the groove of any remnants of the old gasket (use a plastic
scraper and plain water as needed).
3. Line the inside of both sides/walls of the gasket groove with a small
amount of silicon lubricant. This can be sprayed in or brushed in
depending on the type lubricant you use. Make sure to fully coat the
inside of the groove.