Page 15
1.16.
Autoclave Electrical Data
Property
Value
Total Power
2300W
Voltage
1 or 2 phase, / 230VAC
±5%, 60Hz
Amperage
10A
Protection against electrical
shock
IEC 61010-1
Mains supply fluctuation
+/- 5%
Frequency (Hz)
60Hz
Electrical Circuit
Dedicated electrical circuit
Recommended circuit breaker
15A
1.17.
Utilities
Property
Value
Mineral free water
See table in 1.22
Power supply
* 1 or 2 phase, / 230VAC ±5%, 60Hz
Recommended
circuit breaker
15A
* According to the local network.
Cautions!
To avoid any injury by electrical hazard, it is recommended
that a ground fault protection device (GFCI) be installed in
the electrical panel feeding the autoclave (local codes may
make this mandatory).
The electrical network must comply with local rules or
regulations.
Verify that there is an easy access to the main power switch
and to the current leakage safety relay (GFCI). The voltage
supplied to the device must comply with the label ± 5%.
1.18.
Environmental Emission Information
•
The peak sound level generated by the autoclave is 65dBa with
background noise of 48 dBa.
•
The total heat per hour transmitted by the autoclave is <200Wh.
1.19.
Construction
The main parts of the autoclave are made of materials as indicated
below:
•
Chamber is built of stainless steel 316 L.