Page 6
8. Verify once again that you have chosen the appropriate sterilization
program and the water reservoirs are full.
9. Do not place your hand or head, etc. above the door while opening, as
hot steam is escaping the chamber.
10. Do not stand near the back panel of the autoclave while operating, as
the pressure safety valve may release steam.
11. Do not touch hot surfaces, such as the top enclosure and area adjacent
to the chamber opening!
12. After the cycle, open the door slowly to allow steam to escape and wait
1 minute before you remove the load.
13. To assure proper operation of the autoclave, avoid placing in the
vicinity of electrical equipment which is not certified for Electromagnetic
Compatibility according to IEC/EN 61326-1.
14. A certified inspector must perform a periodic pressure chamber safety
test according to the local regulations.
15. Once a year, or more frequently, effectiveness tests must be
performed, i.e., calibration and validation.
16. Make sure there are no leaks, breaks, blockages, whistles, or strange
noises.
17. Perform the maintenance operations as instructed. The owner of the
autoclave is responsible to perform the maintenance operations.
18. Notify the person in charge immediately of any deviation from the
normal functioning of the device.
19. Protective equipment and clothes and other safety instructions should
be implemented in accordance with local and national regulations and/or
rules!