Page 66
Pulse C Stay
Time
Defines the additional
time that the top
exhaust valve will
remain open after
reaching the pressure
set by the Pulse C low
Pressure parameter in
pulse group C.
Seconds
1-100
1
Pulse C Low
Pressure
Defines the pressure
that the top exhaust
valve.
closes at the bottom
of each pulse.
Psia
0.73
–
43.51psia
.01
Pulse C High
Pressure
Defines the pressure
that the top exhaust
valve opens at the top
of each pulse.
Psia
0.73
–
43.51psia
.01
Pulse D Count
Defines the number of
pulses in pulse group
D.
Whole
Number
0-10
1
Pulse D Stay
Time
Defines the additional
time that the top
exhaust valve will
remain open after
reaching the pressure
set by the Pulse D
Low Pressure
parameter in pulse
group D.
Seconds
1-100
1
Pulse D Low
Pressure
Defines the pressure
that the top exhaust
valve.
closes at the bottom
of each pulse.
Psia
0.73
–
43.51psia
.01
Pulse D High
Pressure
Defines the pressure
that the top exhaust
valve opens at the top
of each pulse.
Psia
0.73
–
43.51psia
.01
Sterilization
Temperature
Hea
tin
g
Defines the
temperature that
needs to be reached
during heating to get
to the sterilization
stage.
ºF
212 ºF-
284 ºF
0.5°