Page 97
Caution!
Make sure the gasket sits evenly without waves,
humps, or cavities.
If the door gasket is not perfectly smooth it will not
seal the chamber properly
This drawing shows the correct direction of the
gasket.
Gasket
Door
10.7.
Checking the Safety Valve
There are two safety valves located in the mineral free water reservoir.
To prevent the safety valve from becoming blocked, it is necessary to
open the valve under pressure. This will allow the steam to escape
clearing the seat of the valve. Once per month, perform the following
procedure: