Page 65
steam.
Purge
temperature
Temperature at which
purge stage can start.
°C
Pulse A
Count
Cr
e
a
te
P
u
lse
Defines the number of
pulses in the first
pulse group (group A)
Whole
Number
0-10
1
Pulse A Stay
Time
Defines the additional
time that the top
exhaust valve will
remain open after
reaching the pressure
set by
the Pulse A low
Pressure parameter in
the first pulse group.
Seconds
1-100
1
Pulse A Low
Pressure
Defines the pressure
that the top exhaust
valve
closes at the bottom
of each pulse.
Psia
0.73
–
43.51psia
.01
Pulse A High
Pressure
Defines the pressure
that the top exhaust
valve opens at the top
of each pulse.
Psia
0.73
–
43.51psia
.01
Pulse B
Count
Defines the number of
pulses in pulse group
B.
Whole
Number
0-10
1
Pulse B Stay
Time
Defines the additional
time that the top
exhaust valve will
remain open after
reaching the pressure
set by the Pulse B
Low Pressure
parameter in group B.
Seconds
1-100
1
Pulse B Low
Pressure
Defines the pressure
at which the top
exhaust valve closes
at the bottom of each
pulse.
Psia
0.73
–
43.51psia
.01
Pulse B High
Pressure
Defines the pressure
at which the top
exhaust valve opens
at the top of each
pulse.
Psia
0.73
–
43.51psia
.01
Pulse C
Count
Cr
e
a
te
P
u
lse
Defines the number of
pulses in pulse group
C.
Whole
Number
0-10
1