Page 68
Dry First
Stage Time
The total drying cycle
can be divided into
two stages. This
parameter defines the
length of the first
stage. The second
stage will start at the
end of the first and
last until the end of
the total drying.
cycle.
Minutes
0-120
1
Dry Heat On
2
Defines the ON time
portion of the duty
cycle for the heating
elements during the
second stage of the
drying cycle.
Seconds
0-120
1
Dry Heat Off
2
Defines the OFF-time
portion of the duty
cycle for the heating
elements during the
second stage of the
drying cycle.
Seconds
0-120
1
Additional
Dry Time
Defines the number of
additional minutes to
add to the default dry
time
Minutes
0-60
1
End
Temperature
E
n
d
in
g
Define the
temperature at the
end of the cycle that
must be achieved
before the cycle can
end and the door be
opened.
°F
86°F-
302°F
1
Jacket
Temperature
Glo
b
a
l
This defines the
chamber pre-heat
temperature for the
program selected. It is
not recommended to
go above 120ºF. Too
high a temperature
will damage the
autoclave. The pre-
heat will maintain
the chamber at this
temperature between
cycles and will begin
immediately once the
program is selected
ºF
32- 302ºF
1