100
6.4.2.2
Address 3 (Measurement Mode)
1)
Commands Description
This command is used to set the instrument measurement mode. The instruction data byte bit corresponds
to the function of the instrument as shown in below tables
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
1~32
0x10
0x00
0x03
0x00
0x01
0x01
0x00
0x00or
0x01
Calculated based
on the first 9 bytes
3)
Return Instruction Format
6.4.2.3
Address 4 (Memory Sequence)
1)
Commands Description
This command is used to set the number of the memory sequence in manual mode. Instruction data byte
1 ranges from 1 to 50. The command is invalid when the output is turned on.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data byte
2
CRC
Low
CRC
High
1~32
0x10
0x00
0x04
0x00
0x01
0x01
0x00
0x01~0x32
Calculated based on
the first 9 bytes
3)
Return Instruction Format
0x01
This command is used to set the measure mode to programmable mode. The command is
invalid when the output is turned on.
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x03
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x04
0x00
0x01
Calculated based on the first 6
bytes
Содержание TH7105