135
2)
Send Instruction Format
3)
Return Instruction Format
The return command data byte range is 45.0~500Hz. When the frequency is less than 100 Hz, the
resolution is 0.1 Hz; when the frequency is not less than 100 Hz, the resolution is 1 Hz.
6.4.3.8
Address 8 (Current High Limit)
1)
Command Description
This command is used to query the current high limit value in the current manual mode.
2)
Send Instruction Format
3)
Return Instruction Format
When the set voltage is not greater than 150V, the return command data range is 0.000~8.400A; when
the set voltage is greater than 150V, the return command data range is 0.000~4.200A.
6.4.3.9
Address 9 (Current Low Limit)
1)
Command Description
This command is used to query the current low limit value in the current manual mode.
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x07
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x00
45.0~500, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x08
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x00
0.000~4.200 or
0.000~8.400, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes