6.4.2.20
Address 21 (Surge Drop)
1)
Commands Description
This command is used to control the glitch trap in manual mode. The command is invalid when the output
is turned on. The instruction data byte bit corresponds to the function of the instrument as shown in
below table:
Data Byte
Step Connection
0x00
This command is used to turn off the surge drop in manual mode.
0x01
This command is used to turn on the surge drop in manual mode.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
1~32
0x10
0x00
0x15
0x00
0x01
0x02
0x00
0x00
or
0x01
Calculated based
on the first 9 bytes
3)
Return Instruction Format
6.4.2.21
Address 22 (Over Current Fold)
1)
Commands Description
This command is used to control the over current fold function in manual mode. The command is invalid
when the output is turned on. The instruction data byte bit corresponds to the function of the instrument
as shown in below table:
Data Byte
Over Current Fold Function
0x00
This command is used to turn off the overcurrent fold function in manual mode.
0x01
This command is used to turn on the overcurrent fold function in manual mode.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data
byte 2
CRC
Low
CRC
High
1~32
0x10
0x00
0x14
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x15
0x00
0x01
Calculated based on the first 6
bytes