29
factory settings from M1-1 to M50-9 have the same setting, and each step of the memory needs to be
turned on, after the connection test is completed, the next memory can be executed. For example, in
memory 1, only the step connection of step 1, step 2, step 3, and step 4 is set to "ON", and the step
connection of step 1 of the memory 2 is set to "ON", only M1-1, M1- 2, M1-3, M1-4 are executed,
M2-1 will not be executed, because the step connections of M1-5, M1-6, M1-7, M1-8, M1-9 are “OFF”.
See the programmable mode output function for details.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, the soft-key zone
displays the function menu of this parameter. This parameter can be changed to OFF, Front, or BACK.
AP HiLmt
1)
Description
This parameter is used to set the upper limit of the peak current. When the voltage mode is "AUTO"
and the voltage is less than 150V, the range is 0.0~33.6A; when the voltage mode is "AUTO" and the
voltage is greater than 150V or the voltage mode is "HIGH", the range is 0.0~16.8A. When this
parameter is set to 0.0, this function is turned off.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it with
the front panel numeric keys, and finish with ENTER or OK keys to confirm the change. If you do not
want to change the parameter value or press the wrong numeric key when changing with the numeric
keys, you can cancel it by pressing the
key or ESC key.
AP LoLmt
1)
Description
This parameter is used to set the lower limit of the peak current. When the voltage mode is "AUTO"
and the voltage is less than 150V, the range is 0.0~33.6A; when the voltage mode is "AUTO" and the
voltage is greater than 150V or the voltage mode is "HIGH", the range is 0.0~16.8A. When this
parameter is set to 0.0, this function is turned off.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it with
the front panel numeric keys, and finish with ENTER or OK keys to confirm the change. If you do not
want to change the parameter value or press the wrong numeric key when changing with the numeric
keys, you can cancel it by pressing the
key or ESC key.
P HiLmt
1)
Description
This parameter is used to set the upper limit of the output power. The range is from 0.0 to 1000W.
When this parameter is set to 0.0, this function is turned off.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, then change it with