116
is turned on.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x22
0x00
0x02
0x04
0.000~8.400 or
0.000~4.200, 4
bytes stored in
float point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.34
Address 35 (Setting Frequency)
1)
Command Description
This command is used to set the output frequency value of the current step in the programmed mode.
The command data range is 45.0~500Hz. When the frequency is less than 100 Hz, the resolution is 0.1 Hz;
when the frequency is not less than 100 Hz, the resolution is 1 Hz. Command is invalid when output is
turned on
.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x23
0x00
0x02
0x04
45.5~50, 4 bytes
stored in float
point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.35
Address 36 (Step Connection)
1)
Command Description
This command is used to open the step connection of the current step. Command is invalid when output
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x22
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x23
0x00
0x02
Calculated based on the first 6
bytes