69
Shorthand for command: :FUNC:OUTP OFF or :FUNC:OUTP 0
Command description: This command is used to turn off the AC output
,
this command is only valid under
the measurement display page.
Command Syntax: :FUNCtion:OUTPut?
Shorthand for command: :FUNC:OUTP?
Command description: This command is used to query the current output status, and the command
returns 0 or 1. 0 means the output is off; 1 means the output is on.
6.3.5.2
:FUNC:RM
Command Syntax: :FUNCtion:RunMode:MANUal
Shorthand for command: :FUNC:RM:MANU
Command description: This command is used to set the test mode to manual mode. The command is
invalid when the output is turned on.
Command Syntax: :FUNCtion:RunMode:PROGram
Shorthand for command: :FUNC:RM:PROG
Command description: This command is used to set the test mode to programmable mode. The command
is invalid when the output is turned on.
Command Syntax: :FUNCtion:RunMode?
Shorthand for command: :FUNC:RM?
Command description: This command is used to query the current test mode. The command returns
manual or program. Manual indicates that it is currently in manual mode; program
indicates that it is currently in programmable mode.
6.3.5.3
:FUNC:MEM
Command Syntax: :FUNCtion:MEMory:MANUal
<value>
Shorthand for command: :FUNC:MEM:MANU <value>
Command description: This command is used to set the number of the memory sequence in manual
mode. <value> indicates the number value, and the range is 1~50. The command is
invalid when the output is turned on.
Command Syntax: :FUNCtion:MEMory:MANUal?
Shorthand for command: :FUNC:MEM:MANU?
Command description: This command is used to query the number value of the memory sequence in
manual mode, and returns <value>, its value is equal to 1~50.
Command Syntax: :FUNCtion:MEMory:PROGram
<value>
Shorthand for command: :FUNC:MEM:PROG <value>