79
6.3.5.20
:FUNC:PF:LOLMT
Command Syntax: :FUNCtion:PowerFactor:LOwLiMiT <value>
Shorthand for command: :FUNC:PF:LOLMT <value>
Command description: This command is used to set the power factor low limit value of the current step in
the programmable mode, and <value> indicates the peak power factor low limit
value, which ranges from 0 to 1000W. The command is invalid when the output is
turned on.
Command Syntax: :FUNCtion:PowerFactor:LOwLiMiT?
Shorthand for command: :FUNC:PF:LOLMT?
Command description: This command is used to query the low limit of the power factor of the current
step in the programmable mode, and returns <value>, which is equal to 0~1000W.
The command is invalid when the output is turned on.
6.3.5.21
:FUNC:DELAY
Command Syntax: :FUNCtion:DELAY <value>
Shorthand for command: :FUNC:DELAY <value>
Command description: This command is used to set the delay determination time of the current step in
the programmable mode.<value> indicates the delay determination time. When
the unit is second, the range is 0.1~999.9s; when the unit is minute, the range is
0.1~999.9m; when the unit is time, the range is 0.1~999.9h. The command is
invalid when the output is turned on.
Command Syntax: :FUNCtion:DELAY?
Shorthand for command: :FUNC:DELAY?
Command description: This command is used to query the delay determination time of the current step in
the programmable mode. Returns <value>, when the unit is second, its value is
0.1~999.9s; when the unit is minute, its value is 0.1~999.9m; when the unit is time,
its value is 0.1~999.9h. The command is invalid when the output is turned on.
6.3.5.22
:FUNC:DWELL
Command Syntax: :FUNCtion:DWELL <value>
Shorthand for command: :FUNC:DWELL <value>
Command description: This command is used to set the test time of the current step in the programmable
mode. <value> indicates the test time. When the unit is second, the range is
0.1-999.9s; when the unit is minute, the range is 0.1-999.9m; when the unit is
time, the range is 0.1-999.9h. The command is invalid when the output is turned
on.