This parameter is used to set the status of the display page of the last test result, there are three
modes: “LAST”, “ALL” and “P/F”. When set to "LAST", the display will show the final test result after the
test is over; When set to “ALL”, the display will show all test results after the test is over; When set to
“P/F”, the display will show “PASS” or “FAIL” after the test is over. In the manual mode, when the
voltage output value and the frequency value are in the range of the high voltage limit, the low
voltage limit, the high frequency limit, and the low frequency limit, “PASS” is displayed at the end of
the test; If it is not within the range, “FAIL” is displayed, out of range is displayed, and shows the
parameter values that are out of range. For details, please refer to result display function.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, the soft-key zone
displays the function menu of this parameter, this parameter can be changed by the soft-key LAST,
ALL or P/F.
Surge/Drop
1)
Description
This parameter is used to set whether to enable the Surge/Drop function. When this parameter is set
to “ON”, the MEMORY page will display “SD Volt”, “SD Site”, “SD Time”, “SD Cont”; When this
parameter is set to "OFF", these parameters are not displayed. For details, please refer to Surge/Drop
function.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, the soft-key zone
displays the function menu for this parameter, this parameter can be changed via the soft-key ON or
OFF
OC Fold
1)
Description
This parameter is used to set whether to enable the over-current fold function. When this parameter
is set to "ON", the output voltage will decrease due to the load increased, the current output will be
kept constant at current high limit. (Note: After this function is enabled, the current will output only if
the output current is greater than the current high limit setting value.) When this parameter is set to
"OFF", the over-current fold function is turned off.
2)
Operation
Make use of the
key or the knob to move the cursor to this parameter option, the soft-key zone
displays the function menu of this parameter, this parameter can be changed via the soft-key ON or
OFF.
Dimmer Mode
1)
Description
This parameter is used to set the front dimmer mode or back dimmer mode. When the dimming mode
is turned on, the surge/drop and SD count are enabled by default. In this mode, the manual mode
MEMORY setting can only set parameters such as voltage, frequency, and SD voltage. For details, see