Current Upper Limit(IRMS_H)
1
)
Description
This parameter is used to set current upper limit. When the setting voltage is not greater than 150V,
the range is 0.000~8.400A; When the setting voltage is greater than 150V, the setting range is
0.000~4.200A. When set to 0.000, the current upper limit function is turned off. When output current
is greater than 0.000, there will be an alarm, “HI-A” will be displayed, and output will be turned off.
2
)
Operation
Make use of
key or the knob to move the cursor to this parameter option, change the value via the
front panel numeric keys, then press ENTER or OK to confirm. If you do not want to change the
parameter value or press the wrong numeric key, you can cancel it by pressing the
key or ESC key.
Current lower Limit(IRMS_L)
1
)
Description
This parameter is used to set current lower limit. When the setting voltage is not greater than 150V,
the range is 0.000~8.400A; When the setting voltage is greater than 150V, the setting range is
0.000~4.200A. When the “PLC ONoff” is turned on which is accessed from <SysSetup> then “SysCom”,
and output current is not less than current lower limit, “PASS” will be displayed while the output is
completed.
2
)
Operation
Make use of
key or the knob to move the cursor to this parameter option, change the value via the
front panel numeric keys, then press ENTER or OK to confirm. If you do not want to change the
parameter value or press the wrong numeric key, you can cancel it by pressing the
key or ESC key.
SD Volt
1)
Description
This parameter is used to set surge/drop, and its setting range is 0.0~300V. When the parameter value
is greater than the setting voltage, it is a surge wave, when it is less than the setting voltage, it is a
drop wave. This parameter will be displayed when the Surge/Drop option from COMMON page under
<Setup> page is turned on. For details, please refer to Surge/Drop function instruction.
2)
Operation
Make use of
key or the knob to move the cursor to this parameter option, change the value via the
front panel numeric keys, then press ENTER or OK to confirm. If you do not want to change the
parameter value or press the wrong numeric key, you can cancel it by pressing the
key or ESC key.