163
6.4.3.63
Address 64 (Output Voltage)
1)
Command Description
This command is used to obtain the output voltage. The command is valid when the output is turned on.
2)
Send Instruction Format
3)
Return Instruction Format
Return instruction data range is 0.0~300.0V
6.4.3.64
Address 65 (Output Current)
1)
Command Description
This command is used to obtain the output current. The command is valid when the output is turned on.
2)
Send Instruction Format
3)
Return Instruction Format
Return instruction data range is 0.000~8.400A
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x40
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x04
0.0~300.0, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x41
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x04
0.000~8.400, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes
Содержание TH7105