118
6.4.2.37
Address 38 (AP Low Limit)
1)
Command Description
This command is used to set the peak current low limit value of the current step in the programmed mode.
When the voltage mode is "AUTO" and the voltage is less than 150V, the range is 0.0~33.6A; when the
voltage mode is "AUTO" and the voltage is greater than 150V or the voltage mode is "HIGH", the range is
0.0~16.8A. When this parameter is set to 0.0, this function is turned off. Command is invalid when output
is turned on.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x26
0x00
0x02
0x04
0.0~16.8 or
0.0~33.6, 4 bytes
stored in float
point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.38
Address 39 (Power High Limit)
1)
Command Description
This command is used to set the power high limit of the current step in the programmed mode. The
command data range is from 0.0 to 1000W. Command is invalid when output is turned on
.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x27
0x00
0x02
0x04
0.0~1000, 4 bytes
stored in float
point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
1~32
0x10
0x00
0x25
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x26
0x00
0x02
Calculated based on the first 6
bytes