Replacing the Battery Springs
Replace the individual D cell battery springs when they break or become corroded. Order the battery
spring replacement kit part number 75DK6002-00.
TRDI recommends replacing these springs every five years or whenever visible signs of wear
or corrosion appear.
To replace the battery springs:
1. Remove the batteries.
2. Using a Phillips screwdriver, remove the screw holding the spring in place.
3. Install the new springs.
4. Replace the batteries.
5. Use the
panel to set the date the springs were replaced.
Replacing the Battery Separators
Replace the individual D cell battery separator when they break or become worn. Order the battery sepa-
rator replacement kit part number 75DK6003-00.
To replace the battery separators:
1. Remove the batteries.
2. Remove the separator by turning it counter-clockwise.
3. Install the new separator.
4. Replace the batteries.
Figure 68.