Sentinel V SC and RT Operation Manual
Page 122
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Setting the Date and Time
Setting the Date and Time on Self-Contained Systems
To set the date and time on a Self-Contained system:
to the ADCP.
2. Click the Set Time button to set the Sentinel V ADCP’s clock.
3. On the
Step 4: System check
panel, click the System clock tab.
•
To manually set the time, click on the clock icon. Click on the calendar icon to set the date. Press
the Set Time button.
•
Use the Sync with PC button to set both the time and date of the ADCP to match the time and
date on the PC.
•
Use the Sync with GMT button to set both the time and date of the ADCP to match the GMT
time and date.
4. Click the Close button ( ) to return to the Home panel.
Figure 61.
Set the Date and Time
The real-time clock (date and time) within the Sentinel V is backed up by a super-capacitor
that maintains the correct time while system power is removed. The clock will continue to
maintain the date and time during power outages of 16 to 24 hours. If power is removed for
a longer period of time, then the clock will revert to the default value of January 1, 1970 and
will need to be reset to the correct time.