September
2017
Sentinel V SC and RT Operation Manual
Page 40
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Switching to the Sentinel V Real-Time Mode
(Sentinel V is currently configured for SC mode):
The Sentinel V Self-Contained system MUST have firmware version 47.19.xx or higher before
you can load the Real-Time firmware version 66.xx. If you see the message
“version has to be
greater than 44.17”
even though the firmware version is 47.19.xx, press
CTRL+F5
to continue.
The Sentinel V Self-Contained system MUST have a serial connector on the end-cap and the
user MUST have the 73D-3112-xxx underwater cable and AC power adapter to be able to
switch to the Real-Time mode.
If the system does not have the serial connector, underwater
cable and AC power adapter, once the 66.xx firmware is loaded, there is no easy way to
communicate with the system (see
). Once 66.xx firmware is
installed, the touch sensor is disabled.
Establish a wireless connection and start
ReadyV
.
ReadyV
.
Select
Firmware Upgrade
and then select
the Sentinel V Real-Time firmware binary
(66.xx
.bin
) file. Version 66.xx firmware is
loaded into the system and then the
Sentinel V Real-Time system restarts.
•
Previously installed Real-Time
features are automatically
restored.
•
The wireless connection between
the PC/laptop and Sentinel V Real-
Time is disconnected.
Now
connect the Sentinel V Real-Time
to a
PC/laptop using the serial RS232 cable and
Sentinel V RT Utilities
.