Water Profiling Commands
The following commands define the criteria used to collect the water-profile data.
Available Water Profiling Commands
The available water profiling commands are different for each water mode. The commands listed below
show the commands for each water mode. The added commands are highlighted.
>w?
Available Commands:
WB 0 -------------------- Bandwidth 0=High, 1=low
WC 064 ------------------ Correlation Threshold [0-255]
WD 1111000011 ----------- Data Out {v;c;a;p;s;*;*;*;i;m}
WF 0100 ----------------- Blanking Distance (cm) [0-9999]
WM 0002 ----------------- Water Profiling Mode [2]
WN 030 ------------------ Number of Bins [1-200]
WP 060 ------------------ Number of Pings [0-999]
WS 0200 ----------------- Bin Size (cm) [50-400]
WV 0175 ----------------- Ambiguity Velocity [2-700] (cm/s)
W? ---------------------- Display W-Command Menu
>
>#W?
*** CAUTION: These commands are reserved for RDI use
and may not be currently supported!
Available Commands:
Available Commands:
#WA 050,1 ------------- False Target Threshold 0-255 [, Start bin 0 - 256]
#WB 0 ------------------- Bandwidth 0=High, 1=low
#WE 2000 ---------------- Error Velocity Threshold [0-9999] (mm/s)
#WI 0 ------------------- Bypass phase interpolation (0,1)
#WJ 1 ------------------- BroadBand Rcvr Gain [0=Lo,1=Hi]
#WL 0 0 0 0 -------- Manual dLag Offset <b1> <b2> <b3> <b4> (+/- samples)
#WT 0000 ---------------- Transmit Length (cm)
#W? --------------------- Display #W-Command Menu
>